Corresponding Author

Rong-zong HU

Abstract

A new type of column thin-layer electrolytic selfregenerating suppressor for ion chromatography was proposed.A porous electrode was used in this suppressor,incorporating the electrode and the electrolyte compartment,and eliminating the voltage across the electrolyte compartment.The design of the column thin-layer configuration of suppression compartment simplified the structure of suppressor and reduced the dead volume,improving the performance of suppressor.

Keywords

Suppressor, Ion chromatography, Electrolytic selfregenerating

Publication Date

2006-11-28

Online Available Date

2006-11-28

Revised Date

2006-11-28

Received Date

2006-11-28

References

[1]Tian Z W,Hu R Z,L in H S,et al.H igh-perform anceelectrochem ical suppressor for ion chrom atography[J].J.Chrom atogr.,1988,439:159~163.
[2]L iu K L(刘开录).Continuously self-regenerating h ighperform ance equ ipm ent of ion-exchange[P].CN00201227.8,2000-01.
[3]Pau l R Haddad,Peter E Jackson,M atthew J Shaw.De-velopm ents in suppressor technology for inorgan ic ion a-nalysis by ion chrom atography using conductivity detec-tion[J].J.Chrom atogr.A,2003,1000:725~742.
[4]Saari-Nordhaus Raaidah,Anderson Jr Jam esM.Recentadvances in ion chrom atography suppressor improve an i-on separation and detection[J].J.Chrom atogr.A,2002,956:15~22.
[5]Hu R Z(胡荣宗).The electrochem ical suppressor forion chrom atography w ith incorporation of the electrodeand electrolyte compartm ent[P].CN 02131711.9,2002-09.