Abstract
A MWNTs modified GC electrode was constructed successfully to be used to detect hydrogen peroxide.Two methods of maximum anodic current of cyclic voltommogram and chronoamperometry were adopted to measure the concentration of hydrogen peroxide.The results show that the MWCNTs can enlarge the effective surface area of GC electrode and can enormously enhance the electron transfer.The sensitivity and correlation coefficient for hydrogen peroxide detection are 2.84 μA/mmol·L-1,0.997 and 1.5 μA/mmol·L-1,0.971 for cyclic voltammetry and chronoamperometry,respectively.The procedure of the two methods to detect hydrogen peroxide is simple,and the results obtained are satisfied.
Keywords
MWNTs modified GC electrode, voltammetry, chronoamperometry, hydrogen peroxide, biosensor
Publication Date
2007-02-28
Online Available Date
2007-02-28
Revised Date
2007-02-28
Received Date
2007-02-28
Recommended Citation
Hong-juan WANG, Chun-mei ZHOU, Hao YU, Feng PENG, Jin-kun FANG.
MWNTs Modified GC Electrode for Hydrogen Peroxide Determination[J]. Journal of Electrochemistry,
2007
,
13(1): 72-76.
DOI: 10.61558/2993-074X.1783